https://doi.org/10.1140/epjp/i2017-11646-x
Regular Article
Simulation of Gaussian energy broadening in gamma response of a LYSO array detector using a semi-empirical method
Nuclear Science and Technology Research Institute, Tehran, Iran
* e-mail: atahery@aeoi.org.ir
Received:
7
January
2017
Accepted:
1
August
2017
Published online:
8
September
2017
The energy broadening parameters for a LYSO:Ce array were determined experimentally in an energy range of 59keV to ∼ 1300 keV for gamma rays. The obtained parameters can be used to improve the Monte Carlo simulation of the spectral behavior of the LYSO-based detectors, especially in PET imaging systems. The detector spectral responses were simulated by the MCNP4c code using the obtained parameters. The obtained results from the simulations were verified using real gamma spectra.