Fast transimpedance preamplifier for a boron-coated multiwire proportional chamber neutron detector
1
Key Laboratory of Neutron Physics of China Academy of Engineering Physics, Institute of Nuclear Physics and Chemistry, 621900, Mianyang, China
2
School of Nuclear Science and Technology, University of Science and Technology of China, 230026, Hefei, China
3
Institute of High Energy Physics, Chinese Academy of Sciences, 100049, Beijing, China
* e-mail: Gongjian@hotmail.com
Received:
27
June
2016
Accepted:
30
September
2016
Published online:
26
October
2016
A low-noise and fast preamplifier is needed for the new boron-coated multiwire proportional chamber (MWPC) neutron detector with a delay line readout system to improve position resolution. A transimpedance preamplifier with a rise time of 30ns, a signal-to-noise ratio higher than 40dB, and an automatic gain control function are designed to meet the aforementioned requirements. On the other hand, we also compare it to commercial preamplifiers. Then, the advantages and disadvantages of commercial and in-house preamplifiers are analyzed. The preamplifier is used and tested on a MWPC neutron detector, and results are presented.